Prehospital blood pressures: Inaccuracies caused by ambulance noise?

https://doi.org/10.1016/0735-6757(94)90025-6Get rights and content

Abstract

Blood pressure measurements in a moving ambulance can be difficult to obtain. Sirens, engine noise, and road noise can all interfere with the accurate detection of patient's blood pressure. This study was under-taken to determine the influence of ambulance noise and vibration on auscultated blood pressures. A model was developed that used dynamic pressures to simulate systolic Korotkoff sounds. Forty-nine emergency personnel were asked to obtain blood pressures using the model in both a quiet environment and in a moving ambulance. A total of 485 blood pressure measurements were obtained. Systolic pressures were randomized to two settings: 76 mm Hg and 138 mm Hg. Stationary readings were compared with moving readings using analysis of variance for repeated measures. Systolic blood pressure measurements obtained in the quiet environment averaged 133 ± 5 mm Hg at the high setting, and 45 ± 6 mm Hg at the low setting. Systolic blood pressure measurements obtained in a moving ambulance averaged 86 ± 7 mm Hg at the high setting, and 41 ± 7 mm Hg at the low setting. The average differences between quiet and moving measurements were 47 mm Hg at the “high” setting (P < .01) and 4 mm Hg at the “low” setting (P > .01). At physiological levels, blood pressures obtained in moving ambulances differ significantly from those obtained in a quiet environment, which may be caused by road noise and ambulance motion.