Fluoroscopy-free Resuscitative Endovascular Balloon Occlusion of the Aorta (REBOA) for controlling life threatening postpartum hemorrhage

  • Knut Haakon Stensaeth ,

    knutsten@gmail.com

    Affiliations Dept of Radiology and Nuclear Medicine, St Olavs University Hospital, Trondheim, Norway, Institute of Circulation and Medical Imaging, Norwegian University of Science and Technology, Trondheim, Norway

  • Edmund Sovik,

    Affiliation Dept of Radiology and Nuclear Medicine, St Olavs University Hospital, Trondheim, Norway

  • Ingrid Natasha Ylva Haig,

    Affiliation Dept of Radiology and Nuclear Medicine, Oslo University Hospital Ullevaal, Oslo, Norway

  • Erna Skomedal,

    Affiliation Dept of Radiology and Nuclear Medicine, Stavanger University Hospital, Stavanger, Norway

  • Arve Jorgensen

    Affiliations Dept of Radiology and Nuclear Medicine, St Olavs University Hospital, Trondheim, Norway, Institute of Circulation and Medical Imaging, Norwegian University of Science and Technology, Trondheim, Norway

Abstract

Background

Severe postpartum hemorrhage occurs in 1/1000 women giving birth. This condition is often dramatic and may be life threatening. Resuscitative Endovascular Balloon Occlusion of the Aorta (REBOA) has in recent years been introduced as a novel treatment for hemorrhagic shock. We present a series of fluoroscopy-free REBOA for controlling life threatening postpartum hemorrhage.

Methods

In 2008 an ‘aortic occlusion kit’ was assembled and used in three Norwegian university hospitals. The on-call interventional radiologist (IR) was to be contacted with a response time < 30 minutes in case of life threatening PPH. Demographics and characteristics were noted from the medical records.

Results

This retrospective study includes 36 patients treated with fluoroscopy-free REBOA for controlling severe postpartum hemorrhage in the years 2008–2015. The REBOA success rate was 100% and no patients died from REBOA related complications. Uterine artery embolization was performed in 17 (47%) patients and a hysterectomy in 16 (44%) patients. A short (11cm) introducer length was strongly associated with iliac artery thrombus formation (ρ = 0.50, P = 0.002). In addition, there was a strong negative correlation between uterine artery embolization and hysterectomy (ρ = -0.50, P = 0.002).

Conclusions

Our Norwegian experience indicates the clinical safety and feasibility of REBOA in life threatening PPH. Also, REBOA can be used in an emergency situation without the use of fluoroscopy with a high degree of technical success. It is important that safety implementation of REBOA is established, especially through limited aortic balloon occlusion time and a thorough balloon deflation regime.

Introduction

Hemorrhagic shock is very challenging and is a leading cause of death in trauma patients worldwide [1]. Appropriate management of the massively bleeding patient includes early identification of bleeding sources followed by prompt measures to minimize blood loss, restore tissue perfusion, and achieve hemodynamic stability. Where hemorrhage is controlled expeditiously, patients often recover with little to no morbidity [2].

Severe postpartum hemorrhage (PPH) occurs in 1/1000 women giving birth [3], and is defined by a total blood loos > 1000 ml in the first 24 hours after delivery. This condition is often dramatic and is the leading cause of pregnancy-related death in the USA [2]. Medical approaches to severe PPH include treatments with intravenous fluids and blood products, the use of uterotonic medications such as oxytocin, as well as external uterine massage, curettage and/or an intrauterine balloon. However, the evolution of endovascular techniques has established a role for arterial embolization in the setting of uncontrolled PPH. Embolization of the uterine arteries has the advantages of being a minimally invasive treatment that preserves the uterus without eliminating the possibility of future surgical intervention if necessary [3].

Resuscitative Endovascular Balloon Occlusion of the Aorta (REBOA), which was developed by Edwards et al. in 1953 [4], was initially intended for surgical treatment of abdominal aortic aneurysms and later applied to traumatic hemorrhagic shock [5]. There have been several reports on the use of REBOA in recent years, both for surgical bleeding [6], as well as a temporary measure in Caesarean hysterectomy for placenta percreta [7]. In an emergency situation fluoroscopy is commonly not available, and to achieve immediate control of severe PPH fluoroscopy-free REBOA is therefore often needed. Only a few single case reports on the use of REBOA in PPH have been reported so far [68], in addition to our first report on six patients [9].

Fluoroscopy-free REBOA was introduced in Norway in 2008, and has been performed in three major university hospitals since then. The intention was to introduce a lifesaving procedure with few complications that was easy to perform in an emergency situation. To our knowledge, this is the largest reported series of fluoroscopy-free REBOA to control life threatening PPH.

Materials and methods

Patients and study design

This retrospective study includes all patients, vaginal and caesarean birth (n = 36), treated with fluoroscopy-free REBOA for severe PPH in three major university hospitals in Norway in the years 2008–2015. The contributing hospitals were St. Olavs Trondheim University Hospital (n = 13), Oslo University Hospital, Ullevaal, (n = 13) and Stavanger University Hospital (n = 10). The study was approved by the Regional Ethics Committee, and passive consent was obtained from all included patients.

Aortic balloon occlusion kit

In 2008 an ‘aortic occlusion kit’ was assembled that consisted of different sized PTS-X semi-compliant balloons with diameters ranging from 15 to 30 mm (NuMED Canada Inc., Cornwall, ON, Canada), 6–8 French introducers 11 and 23 cm long (Cordis Corporation, Hialeah, FL, USA), and J-curved fixed core 150 cm 0.035-inch steel guidewires (Boston Scientific, Marlborough, MA, and Argon Medical Devices, Plano, TX, USA). Table 1 shows the main equipment of the kit that was used during the 36 REBOA procedures.

thumbnail
Table 1. Main content of the “aortic occlusion kit” used in all hospitals.

https://doi.org/10.1371/journal.pone.0174520.t001

Intervention

The indication for REBOA was hemodynamically unstable severe PPH, and the on-call interventional radiologist (IR) was to be contacted with a response time < 30 min. All hospitals had a similar REBOA procedure, which has been described previously [9], and all procedures were done by the interventional radiologist on call by using the Seldinger technique through a needle puncture in the common femoral artery under ultrasound guidance due to weak femoral pulses. The balloon catheter was inserted through the introducer over the guidewire approximately 30 cm into the abdominal artery and the balloon was then inflated with isotonic saline (using a 20 ml syringe) until a slight resistance was felt (approximately 7–8 ml for 20 mm balloons and 5–6 ml for 15 mm balloons). With an insertion length of 30 cm the majority of balloons will be placed in aortic zone III (below the renal arteries, but above the aortic bifurcation) [10]. In our routine, we also introduced balloon deflation every 10–15 minutes. The balloon was deflated for approximately one minute before it was re-inflated.

During uterine artery embolization, both arteries were selectively catheterized with a 4 or 5 French catheter, and in most cases medium sized (approximately 2 x 2 mm) gelatin sponge particles were used. In a majority of cases the femoral artery was sealed after removal of the introducer by using the 8 or 6 French Angio-Seal device (St Jude Medical, Minnetonka, MN, USA).

Statistical analyses

Data from all eligible patients were analyzed. Continuous variables are shown as mean ± SD, and ordinal variables as median (range). Selected bivariate relationships were examined with the Spearman rank correlation test. A P value of < 0.05 was considered statistically significant. Analyses were conducted with IBM SPSS Statistics version 23.0.

Results

Demographics and clinical characteristics

Thirty-six patients (age 33.4 ± 5.9) were included in this study. Available demographics and clinical characteristics were obtained from the medical records (Table 2). In all patients, the use of REBOA resulted in an immediate and significant rise in systolic blood pressure (32 ± 22 mmHg), achieving hemodynamic stability. All insertions of the catheter for the occlusion balloon were performed in the OR, however, we have no records of the time from the emergency call to the balloon catheter insertion started.

thumbnail
Table 2. Patient demographics and clinical characteristics available from medical records.

https://doi.org/10.1371/journal.pone.0174520.t002

Six REBOA procedures resulted in complications. Five of them were related to local thrombus formation in the iliac or femoral artery, and all five occurred when using 11 cm introducers. Two of these patients had a surgical thrombectomy and three patients were treated with low-molecular heparin. All five patients recovered successfully. One patient, as previously reported [9], had an aortic tear that was treated surgically and recovered without any permanent sequelae. There were no complications regarding the insertion of the J-curved guide wire or the balloon catheter. The shorter introducer length (11 cm) was strongly associated with arterial thrombus formation (ρ = 0.50, P = 0.002). In addition, there was a strong negative correlation between uterine artery embolization and hysterectomy (ρ = -0.50, P = 0.002).

Discussion

PPH is one of the leading causes of maternal mortality and morbidity.[2] Studies have suggested that many deaths associated with PPH could be prevented with prompt recognition and more rapid and adequate treatment [11,12]. Several studies have also noted an increase in PPH in high-resource countries, including Australia and Norway [13,14]. The use of expensive blood products may also be reduced by the use of REBOA, this is especially important in low-resource countries. The current study is the first series with systematic analysis on the subject of fluoroscopy-free REBOA for controlling life threatening PPH. REBOA elevated SBP in the setting of hemorrhagic shock in severe PPH and gained time for further evaluation and treatment, such as uterine artery embolization [15]. REBOA appears to be a quite safe procedure, and in our series no patients died from REBOA-related complications.

The main causes of PPH include uterine atony after childbirth, placenta adhesion, soft birth canal laceration, and coagulation abnormalities, the first by far being the most common reason [16]. Hysterectomy is an effective method to stop PPH, however, in combination with REBOA there is more time to wait for the effects of drugs given to induce contraction of the atonic uterus. If bleeding continues, uterine artery embolization should be considered. In combination with REBOA, this interventional technique can rapidly stop bleeding and preserve the reproductive function of the patient. This fact explains the correlation between uterine artery embolization and hysterectomy in our series.

There are several reports of using fluoroscopy for balloon positioning, however, fluoroscopy is time consuming and rarely available in the obstetric OR and settings with limited medical infrastructure. Guilani et al. have even shown that ultrasonography alone is safe and accurate for positioning and deployment of the balloon catheter [17]. Also, appropriately trained medical personnel with sufficient knowledge of the risk factors and procedures associated with an endovascular approach (e.g. interventional radiologists, anesthesiologists, and vascular surgeons) are often available during emergency situations. A recent review showed that from 16 fluoroscopy-free REBOA studies there was only one serious placement-related complication (aortic tear) [9], and precautions have now been taken to avoid this issue through reduction of the balloon size. Therefore, based on the low complication rate in previous and the present study, we recommend balloon positioning without the use of fluoroscopy.

In trauma patients there are several studies that have reported no vessel injuries caused by the femoral artery introducer or balloon catheter [18,19], however, a study by Saito et al. included two major complications with lower limb ischemia [20], 10 French introducers were used in these cases, in contrast to the present study where 6–8 French introducers were used. Complications in the present study were related to local iliac artery thrombus formation, which was significantly correlated to shorter introducer lengths (11 vs. 23 cm). Due to the sinuous iliac arteries it is possible that short introducers may affect the iliac vessel wall and thereby facilitate thrombus formation, while longer introducer lengths, more often will reach the abdominal aorta.

In the present study, no complications related to balloon occlusion time were observed, however, occlusion time was limited (33.3 ± 5.9 min.). In experiments with swine, Morrison et al. reported that a longer inflation time induces an inflammatory response through the release of interleukin 6, increases the incidence of adult respiratory distress syndrome, and increases the use of vasopressors [21]. A recent study showed that a REBOA duration of 60 min. in a swine model was well tolerated [22]. Furthermore, it is possible that the balloon sometimes ends up in the aortic zone II with the potential of occluding visceral and/or renal arteries. In a study of laparoscopic partial nephrectomy, a warm ischemia time of 40 min. was well tolerated before a significant decrease in renal function occurred [23].

Short balloon deflations allow reperfusion of the extremities and thereby reduces distal ischemia and the risk of thrombosis in non-heparinized patients. In the present study, repetitive deflations every 10–15 min. (each < 1 minute duration) were performed with an immediate deflation and balloon retraction after hemorrhage control. It is reasonable to believe that a combination of a short total occlusion time and repetitive balloon deflations probably will facilitate a minimum of ischemic complications. In addition, the balloon deflations gave time to evaluate if the patient was still bleeding. In a porcine hemorrhagic shock model, Russo et al. showed that partial balloon occlusion of the aorta (pREBOA) resulted in more physiologically tolerable hemodynamic and ischemic changes compared with complete occlusion [24]. Therefore, we believe it is important with a standardized procedure in REBOA to include either a balloon deflation regime or pREBOA.

The development of new devices, which do not require an oversized introducer, is likely to reduce not only complications but also time to occlusion [25]. In the last couple of years 7 French balloon catheters and introducers have been clinically available. Recently in Norway, a 15 mm diameter PTS-X balloon has been used in a few REBOA cases, which only require a 6 French introducer. Our experience is that in a hypovolemic PPH patient, a 15 mm balloon is sufficient to occlude or partially occlude the infra-renal aorta [26], and in this way stabilize the hemodynamic emergency situation.

Limitations

Limitations in this study include particularly the small number of evaluated patients (n = 36), and that it is based on a retrospectively applied definition. However, to our knowledge, this is the largest study so far. Furthermore, a randomized controlled trial and/or the use of propensity scores is difficult to perform in the acute care setting. Additional multicenter studies are required to determine the effectiveness of REBOA.

Conclusions

The present study indicates that fluoroscopy-free REBOA in an emergency situation can be performed with a high degree of technical success, to achieve hemodynamic stability, and to reduce mortality and morbidity associated with severe PPH.

During severe PPH, the combination of REBOA with uterine artery embolization should be considered to prevent unnecessary hysterectomies and to preserve the reproductive function of the patient.

Since adverse effects of invasive procedures cannot be avoided, the acquisition of standard procedures and safety implementation of REBOA is important.

Author Contributions

  1. Conceptualization: KHS E. Sovik.
  2. Formal analysis: AJ.
  3. Investigation: KHS AJ.
  4. Methodology: KHS E. Sovik.
  5. Project administration: KHS.
  6. Resources: AJ YH E. Skomedal.
  7. Supervision: KHS.
  8. Validation: KHS.
  9. Visualization: KHS E. Sovik AJ.
  10. Writing – original draft: KHS.
  11. Writing – review & editing: KHS E. Sovik AJ.

References

  1. 1. Krug EG, Sharma GK, Lozano R. The global burden of injuries. Am J Public Health 2000 90: 523–526. pmid:10754963
  2. 2. Chichakli LO, Atrash HK, MacKay AP, Musani AS, Berg CJ. Pregnancy-related mortality in the United States due to hemorrhage: 1979–1992. Obstet Gynecol 1999 94: 721–725. pmid:10546717
  3. 3. Lee JS, Shepherd SM. Endovascular treatment of postpartum hemorrhage. Clin Obstet Gynecol 2010 53: 209–218. pmid:20142657
  4. 4. Edwards WS, Salter PP, Carnaggio VA. Intraluminal aortic occlusion as a possible mechanism for controlling massive intra-abdominal hemorrhage. Surg Forum 1953 4: 496–499. pmid:13187327
  5. 5. Gupta BK, Khaneja SC, Flores L, Eastlick L, Longmore W, Shaftan GW. The role of intra-aortic balloon occlusion in penetrating abdominal trauma. J Trauma 1989 29: 861–865. pmid:2661845
  6. 6. Harma M, Kunt AS, Andac MH, Demir N. Balloon occlusion of the descending aorta in the treatment of severe post-partum haemorrhage. Aust N Z J Obstet Gynaecol 2004 44: 170–171. pmid:15089847
  7. 7. Paull JD, Smith J, Williams L, Davison G, Devine T, Holt M. Balloon occlusion of the abdominal aorta during caesarean hysterectomy for placenta percreta. Anaesth Intensive Care 1995 23: 731–734. pmid:8669612
  8. 8. Masamoto H, Uehara H, Gibo M, Okubo E, Sakumoto K, Aoki Y. Elective use of aortic balloon occlusion in cesarean hysterectomy for placenta previa percreta. Gynecol Obstet Invest 2009 67: 92–95. pmid:18936549
  9. 9. Søvik E, Stokkeland P, Storm BS, Asheim P, Bolås O. The use of aortic occlusion balloon catheter without fluoroscopy for life-threatening post-partum haemorrhage. Acta Anaesthesiol Scand 2012 56: 388–393. pmid:22260088
  10. 10. Stannard A, Eliason JL, Rasmussen TE. Resuscitative endovascular balloon occlusion of the aorta (REBOA) as an adjunct for hemorrhagic shock. J Trauma 2011 71: 1869–1872. pmid:22182896
  11. 11. Berg CJ, Harper MA, Atkinson SM, Bell EA, Brown HL, Hage ML, et al. Preventability of pregnancy-related deaths: results of a state-wide review. Obstet Gynecol 2005 106: 1228–1234. pmid:16319245
  12. 12. Della Torre M, Kilpatrick SJ, Hibbard JU, Simonson L, Scott S, Koch A, et al. Assessing preventability for obstetric hemorrhage. Am J Perinatol 2011 28: 753–760. pmid:21698554
  13. 13. Ford JB, Roberts CL, Simpson JM, Vaughan J, Cameron CA. Increased postpartum hemorrhage rates in Australia. Int J Gynaecol Obstet 2007 98: 237–243. pmid:17482190
  14. 14. Rossen J, Okland I, Nilsen OB, Eggebø TM. Is there an increase of postpartum hemorrhage, and is severe hemorrhage associated with more frequent use of obstetric interventions. Acta Obstet Gynecol Scand 2010 89: 1248–1255. pmid:20809871
  15. 15. Xu JQ. Effectiveness of embolization of the internal iliac or uterine arteries in the treatment of massive obstetrical and gynecological hemorrhages. Eur Rev Med Pharmacol Sci 2015 19: 372–374. pmid:25720705
  16. 16. Carroli G, Cuesta C, Abalos E, Gulmezoglu AM. Epidemiology of postpartum haemorrhage: a systematic review. Best Pract Res Clin Obstet Gynaecol 2008 22: 999–1012. pmid:18819848
  17. 17. Guliani S, Amendola M, Strife B, Morano G, Elbich J, Albuquerque F, et al. Central aortic wire confirmation for emergent endovascular procedures: As fast as surgeon-performed ultrasound. J Trauma Acute Care Surg 2015 79: 549–554. pmid:26402527
  18. 18. Brenner ML, Moore LJ, DuBose JJ, Tyson GH, McNutt MK, Albarado RP, et al. A clinical series of resuscitative endovascular balloon occlusion of the aorta for hemorrhage control and resuscitation. J Trauma Acute Care Surg 2013 75: 506–511. pmid:24089121
  19. 19. Ogura T, Lefor AT, Nakano M, Izawa Y, Morita H. Nonoperative management of hemodynamically unstable abdominal trauma patients with angioembolization and resuscitative endovascular balloon occlusion of the aorta. J Trauma Acute Care Surg 2015 78: 132–135. pmid:25539214
  20. 20. Saito N, Matsumoto H, Yagi T, Hara Y, Hayashida K, Motomura T, et al. Evaluation of the safety and feasibility of resuscitative endovascular balloon occlusion of the aorta. J Trauma Acute Care Surg 2015 78: 897–903; discussion 904. pmid:25909407
  21. 21. Morrison JJ, Ross JD, Markov NP, Scott DJ, Spencer JR, Rasmussen TE. The inflammatory sequelae of aortic balloon occlusion in hemorrhagic shock. J Surg Res 2014 191: 423–431. pmid:24836421
  22. 22. Scott DJ, Eliason JL, Villamaria C, Morrison JJ, Houston R, Spencer JR, et al. A novel fluoroscopy-free, resuscitative endovascular aortic balloon occlusion system in a model of hemorrhagic shock. J Trauma Acute Care Surg 2013 75: 122–128. pmid:23940855
  23. 23. Godoy G, Ramanathan V, Kanofsky JA, O’Malley RL, Tareen BU, Taneja BU, et al. Effect of warm ischemia time during laparoscopic partial nephrectomy on early postoperative glomerular filtration rate. J Urol 2009 181: 2438–43; discussion 2443. pmid:19371905
  24. 24. Russo RM, Neff LP, Lamb CM, Cannon JW, Galante JM, Clement NF, et al. Partial Resuscitative Endovascular Balloon Occlusion of the Aorta in Swine Model of Hemorrhagic Shock. J Am Coll Surg 2016 223: 359–368. pmid:27138649
  25. 25. Teeter WA, Matsumoto J, Idoguchi K, Kon Y, Orita T, Funabiki T, et al. Smaller introducer sheaths for REBOA may be associated with fewer complications. J Trauma Acute Care Surg 2016
  26. 26. Davis AE, Lewandowski AJ, Holloway CJ, Ntusi NA, Banerjee R, Nethononda R, et al. Observational study of regional aortic size referenced to body size: production of a cardiovascular magnetic resonance nomogram. J Cardiovasc Magn Reson 2014 16: 9. pmid:24447690