The use of the Cortrak Enteral Access System™ for post-pyloric (PP) feeding tube placement in a Burns Intensive Care Unit

Burns. 2011 Mar;37(2):277-80. doi: 10.1016/j.burns.2010.07.011. Epub 2010 Nov 5.

Abstract

Introduction: 50% of critically ill patients fail to reach caloric targets with NG feeding. PP feeding may enhance caloric intake. PP feeding can be continued throughout theatre in patients with a secure airway. Blind PP tube placement is difficult. CEAS has been developed to assist tube placement and eliminate check X-rays of tube position.

Method: All BITU patients with CEAS PP feeding tube placement were identified. Notes and X-rays were reviewed. Tube position, calorie deficit and time off feed were recorded.

Results: 44 tubes were placed in 21 patients using CEAS. 84% were PP, 16% NG. Position correlated to X-ray findings in 86%. In 16% position was NG on CEAS but was PP on X-ray. 10 patients required both CXR and AXR to confirm position, the remainder required CXR only. Time off feed varied from 0-24 h (mean 7.4 h). Calorie deficit ranged from 0-2465 kCal (mean 858 kCal). Average wait for X-ray was 3.4h. If X-ray wait was eliminated calorie deficit would be reduced by 45% to 393 kCal.

Conclusion: The Cortrak system is safe and effective on BITU. It reduces calorie deficit, reduces X-ray exposure and is cost effective. We recommend its use on BITU.

MeSH terms

  • Burns / economics
  • Burns / therapy*
  • Critical Care*
  • Energy Intake
  • Enteral Nutrition / economics
  • Enteral Nutrition / instrumentation*
  • Humans