Improved diagnostic value of a TCD-based non-invasive ICP measurement method compared with the sonographic ONSD method for detecting elevated intracranial pressure

Neurol Res. 2014 Jul;36(7):607-14. doi: 10.1179/1743132813Y.0000000308. Epub 2014 Jan 12.

Abstract

Objectives: To compare the diagnostic reliability of optic nerve sheath diameter (ONSD) ultrasonography with a transcranial Doppler (TCD)-based absolute intracranial pressure (ICP) value measurement method for detection of elevated ICP in neurological patients. The ONSD method has been only tested previously on neurosurgical patients.

Methods: A prospective clinical study of a non-invasive ICP estimation method based on ONSD correlation with ICP and an absolute ICP value measurement method based on a two-depth TCD technology has recruited 108 neurological patients. Ninety-two of these patients have been enrolled in the final analysis of the diagnostic reliability of ONSD ultrasonography and 85 patients using the absolute ICP value measurement method. All non-invasive ICP measurements were compared with 'Gold Standard' invasive cerebrospinal fluid (CSF) pressure measurements obtained by lumbar puncture. Receiver-operating characteristic (ROC) analysis has been used to investigate the diagnostic value of these two methods.

Results: The diagnostic sensitivity, specificity, and the area under the ROC curve (AUC) of the ONSD method for detecting elevated intracranial pressure (ICP >14·7 mmHg) were calculated using a cutoff point of ONSD at 5·0 mm and found to be 37·0%, 58·5%, and 0·57, respectively. The diagnostic sensitivity, specificity, and AUC for the non-invasive absolute ICP measurement method were calculated at the same ICP cutoff point of 14·7 mmHg and were determined to be 68·0%, 84·3%, and 0·87, respectively.

Conclusions: The non-invasive ICP measurement method based on two-depth TCD technology has a better diagnostic reliability on neurological patients than the ONSD method when expressed by the sensitivity and specificity for detecting elevated ICP >14·7 mmHg.

Keywords: Absolute intracranial pressure,; Diagnostic reliability,; Doppler technology,; Non-invasive measurement,; Optic nerve sheath diameter.

Publication types

  • Clinical Trial
  • Comparative Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Adolescent
  • Adult
  • Aged
  • Aged, 80 and over
  • Area Under Curve
  • Female
  • Humans
  • Intracranial Hypertension / cerebrospinal fluid
  • Intracranial Hypertension / diagnosis*
  • Intracranial Hypertension / diagnostic imaging*
  • Intracranial Pressure* / physiology
  • Male
  • Middle Aged
  • Optic Nerve / diagnostic imaging*
  • Prospective Studies
  • ROC Curve
  • Reproducibility of Results
  • Sensitivity and Specificity
  • Spinal Puncture
  • Ultrasonography / methods*
  • Ultrasonography, Doppler, Transcranial / methods*
  • Young Adult